Here you can find the NEWSLETTER archive More ...
HJM PMS ClearClean Vaisala

cleanroom online
Deutsch   English


  • Particles

Highly Accurate Low Differential Pressure Sensor

The EE610 is ideal for reliable and precise measurement of low differential pressure. It features selectable ranges for ±25, ±50, ±100 and 0...100 Pa.

EE610 Differenzdrucksensor für niedrige Druckbereiche. (Foto: E+E Elektronik Ges.m.b.H.) / EE610 differential pressure sensor for low pressure. (Photo: E+E Elektronik GmbH)
EE610 Differenzdrucksensor für niedrige Druckbereiche. (Foto: E+E Elektronik Ges.m.b.H.) / EE610 differential pressure sensor for low pressure. (Photo: E+E Elektronik GmbH)

With the EE610, E+E Elektronik offers a new sensor for low differential pressure up to ±100 Pa with an accuracy of ±0.5 Pa. Due to selectable measuring ranges and versatile configuration options, the EE610 is suitable for a variety of applications in air as well as in non-flammable and non-aggressive gases.

The EE610 is dedicated for medical, pharmaceutical and clean room applications. For instance, the sensor can measure the differential pressure between clean rooms, surgery rooms or isolation chambers. The non-flow through design excludes cross-contamination.

Selectable Measuring Ranges

The EE610 enables unidirectional or bidirectional differential pressure measurement within a selectable measuring range of 0...100 Pa or ±25, ±50 and ±100 Pa. The long-term stable, piezo-resistive pressure sensing element provides a high accuracy of ±0.5 Pa

Analogue Output and Display

The measured data is simultaneously available as current and voltage signals at the spring loaded terminals. The illuminated graphic display shows the reading in Pa, mbar, inch water column or mm H2O.

Fully Configurable

The EE610 can be easily configured on site. Pressure range, time constant, output signal, display units and backlight can be set via DIP switches on the electronics board. The zero and span point adjustment can be performed with push buttons.

Easy Mounting

The functional IP65 / NEMA 4 enclosure facilitates mounting the sensor with closed cover. This saves time and also protects the electronics from construction site pollution and mechanical damage during installation. For wiring the EE610 features both a cable gland and a knockout opening for a ½”conduit adapter.



E+E_Logo_100px_72dpi_RGB_LI
E+E Elektronik Ges.m.b.H.
Langwiesen 7
4209 Engerwitzdorf
Austria
Phone: +43 7235 6050
Fax: +43 7235 6058
email: info@epluse.com
Internet: http://www.epluse.com


Better informed: With YEARBOOK, NEWSLETTER, NEWSFLASH and EXPERT DIRECTORY

Stay up to date and subscribe to our monthly eMail-NEWSLETTER and our NEWSFLASH. Get additional information about what is happening in the world of cleanrooms with our printed YEARBOOK. And find out who the cleanroom EXPERTS are with our directory.

C-Tec Becker PPS MT-Messtechnik