- Building technology
Air outlet particle filter for the MAS-100 NT microbial air sampler
The MAS-100 NT microbial air sampler from MBV AG is specified for use in ISO Class 5 and GMP Grade A clean rooms. Construction, surfaces and components are of the highest quality and do not produce significant amounts of particles that contaminate the environment.
However, there are situations where a user wants to use his air sampler in different cleanliness classes. This creates a low risk of contamination within a device being carried-over from one environment to another and then blown out.
The newly available optional air outlet filter for the MAS-100 NT air sampler has the following properties:
- HEPA H13 filter with more than 99.95% separation rate for particles > 0.3µm.
- No unfiltered dead volume can be transferred as the filter is positioned directly at the air outlet.
- The filter adapter can be retrofitted by the user for all MAS-100 NT. Recalibration of the airflow sensor is recommended.
- Filters can be exchanged by the user. No recalibration is required.
- Filters are specified for at least one year of use.
The filter module is available immediately in the following configurations:
- MAS-100 NT with filter adapter and filter
- Retrofit set incl. filter for all MAS-100 NT
- Replacement filter
For more information, including data on the separation rate, the long-term behaviour and the transfer of parti-cles between cleanroom classes, please visit www.mbv.ch.
Contact: Roland Durner, roland.durner@mbv.ch, +41 44 928 30 82
MBV AG
Microbiology and Bioanalytic
Industriestrasse 9
8712 Stäfa
Switzerland
Phone: +41 44 9283080
email: welcome@mbv.ch
Internet: http://www.mbv.ch