Kalbrierung of smallest pressures in the wind tunnel
The today usual mechanisms for the representation of small pressures
are based on the principle of a kolbenmanometers or a sealing cone.
With these procedures small pressures can be represented starting from
1 Pa reproducibly. The messunsicherheit amounts to +/- the PTB for
some time looks for 0.3 Pa for a more exact procedure for the
representation of small pressures, in order to become fair the
requirements of the industry. One of the substantial demands for pure
areas is that the positive pressure to the ambient pressure approx..
15 Pa amounts to. Around these pressure values to guarantee most exact
measuring instruments are necessary. The used measuring instruments
must be examined regularly in the area of application between 1 and 20
Pa. There however an uncertainty of 0,3 Pa with itself are only
conditionally suitable they bring the calibration with the past
calibration methods for the calibration of the measuring instruments
used in the pure area.With a new, patented measuring procedure in the wind tunnel Testo can
become fair industrial services the demands of the ISO 14644 or VDI
2083. These standards must be kept in the producing industry, e.g.
Pharma, biotechnology and electro-technology and be fulfilled with
start-up of the pure areas by the manufacturers. With the help of this
calibration procedure gauges within a range can be kalbriert smaller
than 20 Pa DKD pressure starting from 0,1 Pa to be reproducibly
represented there can. Besides the smallest definable messunsicherheit
is appropriate in the range for smaller 20 Pa with +/- 0.02 Pa under
that of the past procedures... far information find you clearly in the
pdf document